Description

Salt spray test machine – DHSP-160

Structure

1.With imported high-precision special intelligent temperature humidity controller, keep effective control of temperature, humidity and dry.
2.The two intake pressure, ensuring stability.
3.A spray tower with conical diffuser, salt fog guiding and salt spray uniformity and the settlement adjusting.
4.Perfect protection function (dry, wet, overload etc.) and devices with voice alarm, stop function.
5.The compressed air is saturated with spray treatment on the test, to ensure stable environmental conditions, effectively prevent the nozzle blocking.
6.With running time controller, after finishing accumulating will shut down automatically.

Protection

(1) Automatically cut off the power supply when the water level is low
(2) Super temperature, automatic cut-off power supply device
(3) With safety warning lamp device
(4) Leakage with automatic shutdown and other protection

See Other Products

Customize Chamber
Temperature and Climatic Test Chambers in Combination with Vibration

Technical Specifications

1. Product Name
Model
Salt spray test machine
DHSP-160
2Sample Forbidden Forbidden for the Test chamber:
Test or stock of flammable, explosive, volatile test samples
Test or stock of corrosive test samples
Test or stock of organisms
Test or stock of o Strong electromagnetic radiations sources
For Test or stock of special test, please contact with us.
3Volume, dimension and weight
3.1 Internal volume 880L
3.2 Working room dimension L1600mm*W1000mm*H550mm
3.3 External dimension L2600mm*W1450mm*H1550mm
3.4 Net Weight About 600kg
4Performance
4.1 Environment Temperature is +25℃,relative humidity ≤85%, Free load.
4.2 Temperature range Test room: 15~70℃
4.3 Humidity range 30~98%R.H
4.4 Temperature fluctuation: ±0.5℃
4.5 Temperature uniformity ≤2.0℃
4.6 Salt spray deposition 1-2ml/h.80cm²
4.7Spray Method Continuous salt spray (ASTM standard) or periodic salt spray (GB standard)
4.8Power Voltage 1∮ AC220V±10% 60HZ
4.9Protection device With double over temperature protection, water shortage warning to ensure safe use
5Product Instruction:
5.1 Structure

1.With imported high-precision special intelligent temperature humidity controller,   keep effective control of temperature, humidity and dry.
2.The two intake pressure, ensuring stability.
3.A spray tower with conical diffuser, salt fog guiding and salt spray uniformity and the settlement adjusting.
4.Perfect protection function (dry, wet, overload etc.) and devices with voice alarm, stop function.
5.The compressed air is saturated with spray treatment on the test, to ensure stable environmental conditions, effectively prevent the nozzle blocking.
6.With running time controller, after finishing accumulating will shut down automatically.

5.2 Material Structure 1.Test tank: high temperature resistant HC PVC plastic plate of 6mm;
2. Test machine cover: High temperature resistant PVC HC plastic plate 10mm;
A  cover for Herringbone arch shape (Condensed water will return  back to machine from inclined wall to avoid the phenomenon of reflux)
B   Cover and the machine are sealed by water to prevent the leakage of the salt spray;
C  Cover is opened for the pneumatic cylinder automatic switch, operation is convenient;
3 Machine shell: Imported PVC board 8mm, the middle layer with stainless steel frame to strengthen the overall solid and carrying capacity of the test machine, in the middle of foam for insulation;
4 Saturated air tank and function: 3mm thick SUS#304 stainless steel plate, and attached to the water level indicating device for the heating ,humidification and oil filter of compress the air into the machine.
5. The test sample frame can freely adjust the test angle 15 degrees to 30 degrees, at an angle of 30 degrees specimen will be up to 50 pieces at least.
5.3
According to CNS.JIS.ASTM IEC Standard
1 Salt spray test: NSS,ACSS
(1) Test room: 35℃±1℃
(2) Saturated air barrel: 47℃±1℃
2 Corrosion resistance test: CASS
(1) Test room: 50℃±1℃
(2) Saturated air barrel: 63℃±1℃
3.Dry test
(1) Test room: 40~70℃±0.5℃
(2) Humidity: 30%-98±2%RH
4.Constant temperature and humidity test:
Test temperature:10℃-70℃±0.5
Humidity:30%-98±2%RH
GB2423.393 Damp heat test :<Basic environmental testing procedures for electric and electronic products > test Ca: Damp heat test methods. (constant temperature and humidity test just meet the specified here)
1, Test temperature: 43 ℃ + 1.
2,Test humidity: 85%~98% + 2%RH.
3,By the use of steam, through the heat water to produce steam, and then into the working room. With the control of microcomputer processor, when the humidity does not meet the need of the value, is inching control; when the humidity exceeds the value, the dry hot air come into, will dilute the moisture and the excess the moisture is discharged. The humidification system using electronic and micro way humidification system (SSR driver).
A, Static room temperature experiment:
The cooling time is less than 15 minutes.
The external air is introduced into the cooling chamber so that the temperature of the test chamber is consistent with the outdoor temperature.
In the standard atmospheric pressure to do the static room temperature experiment, by the conventional spray test to the static placement experiment, mainly through two ways of cooling:
First, because the first step is the conventional spray test, so the test room temperature: 35 degrees, down to room temperature is at room temperature is used to blow air blower down to room temperature. The effect of static experiment second: if placed in a temperature, you specify, for example: 23 degrees in summer when it is difficult to achieve, you need to add a refrigeration system to achieve the purpose of cooling.
All of the test steps are programmed by the controller before the system has been set up, and achieve convergence.
5.4 Air supply system The air pressure is 1kg/cm – two sections adjustment
(1)First section of 2-2.5 kg/cm simple adjustment with filtered oil water function
(2) Second 1 kg/cm 1/4PT precision adjustment pressure meter
5.5 Spray mode: 1.By the principle of “Panut” absorbing saline and atomization, well uniform atomization , without blocking crystallization, can ensure the standard of continuous testing.
2.Nozzle: made of special glass, can adjust the size and quantity of spray and angle.
3.Spray volume :Adjustable: 1-2ml/80cm – /hr (16 hour average)
4.Mist removing method: Automatic mist removing device
5.Principle: Gas generated by the air compressor, enter into the working room, excess fog will be dispersed, exhaust gas discharged through the holes on the back to ensure the cover without fog and condensation water, Corrosion condition can be clearly observed.
5.6 Heating method Corrosion-resistant titanium alloy high speed electric heater, heating fast to reduce the standby time, when the temperature reaches as set, ON, OFF, switch, accurate temperature, less power consumption.
5.7 Control system: Temperature and humidity controller: Touch screen programmable controller.
1, LCD display, according to health settings.
2, with the fixed point and the program executive function, hot and cold boot slope setting operation parameters can provide a choice
3, Chinese screen display, with RS232 communication interface.
B, no fuse power switch: 380V three-phase 100A.
C, the operation time timer: 0~999hr (1 groups).
D, operation switch and indicator light (1 groups).
E, the security alarm control system: low water level, super temperature, overload, etc., additional warning lights.
5.8Refrigeration System 1, Fully enclosed refrigeration compressor
2, The whole system uses R13 environmental refrigerant
3, Heating, cooling system is completely independent
4, The laboratory use of titanium alloy evaporator to prevent corrosion
5.9Heating System 1, Using high speed titanium alloy electric heater
2, Completely independent system, does not affect the freezing and control lines
5.10               Humidity system 1, Humidification, desiccant system requirements are completely independent
2, The water level control adopts mechanical floating ball control, to eliminate electronic false action
3, The water supply system using automatic replenishment system, suitable for long time continuous operation of the machine.
6.Safety Protection Device
Protection (1) Automatically cut off the power supply when the water level is low
(2) Super temperature, automatic cut-off power supply device
(3) With safety warning lamp device
(4) Leakage with automatic shutdown and other protection
7. Consumables
Normal spare parts(Optional) Glass nozzle, sodium chloride, paper, glass filters, distilled water, copper chloride
8 Transportation
8.1 Transportation dimension L2800mm*W1650mm*H1800mm(Approx.)
8.2 Transportation Weight 750kg
9. Use Conditions: Supplied by user
9.1 Installation site Surface smooth, well-ventilated
No strong vibration around
No strong magnetic fields affect
Non-flammable, explosive, corrosive substances and dust
Installation requirements: Around the machine with certain space for use and maintenance (200 long *300 wide cm).
Installation near to corner or on the bottom with a hole diameter of about 10cm for gas outlet or drainage to outside.
9.2 Environment Conditions Temperature: 5℃–35℃,Relative humidity: ≤85%,   pressure: 86kPa—106 k Pa
9.3 Power supply conditions Range of voltage fluctuation: AC(220±10%)V 15A;
Requires the user to install the appropriate air or power switch, and it is better for machine only.
Range of frequency Fluctuations : (50±0.5)HZ
Grounding resistance of Ground Protection is less than 4 amps
User should provide corresponding volume of air or power switch in installation site and the switch should be independent for the use of this machine.
9.4 Others 1, Users need to prepare gas source or air compressor
2, Compressed air: more than 3KG.
3, Saline solution requires pure water or distilled water solution
9.5 Power capacity 10KW
9.6 Max current 20A